Single Use Loop
TruFluor DO
![]() Features
|
TruFluor DO
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Environmental Specifications | ||||
|---|---|---|---|---|
| Parameter | Units | Min | Type | Max |
| Operating Temperature |
°C | 4 | 45 | |
| Pressure | Bar (psi) | (0 non- freezing) |
1 (14.7) |
2 (29.4) |
| Storage Temperature | °C | 0 | 22 | 65 |
| Relative Humidity (non-condensing) |
% | 0 | 60 | 100 |
| Altitude | feet | 40k | ||
| Materials | USP Class VI and 316L Electro-polished SS | |||
| Performance Specifications | ||
|---|---|---|
| |
Min | Max |
| Sample Rate | Sample rate can be set so that samples are taken in increments from 1 second to 30 minutes | |
| Measuring Range O2 | ~ 0 | 52.5% O2 ( 250% air-sat) |
| Limit of Detection | 0.03% O2 | |
| Precision | 0.55% or < 3% (whichever is greater) of reading at all temperatures and O2 levels < 21% * | |
| Accuracy at 20 °C | < ± 1% at 20.95% O2 |
|
| Response Time (sec) | <40 (agitated, 90%) | <60 (not agitated, 90%) |
| Calibration | Pre-calibrated (gamma-radiation-resistant chip) | |
| Gamma Radiation | Tested to 50 kGy (typical 25 - 40 kGy) | |
| Cross Sensitivity | Organic solvents (e.g. toluene, acetone, benzene, methylene chloride) SO2 and Cl2 gas | |
| Temperature | Accuracy | Precision |
| From 5 to 45°C ** | ± 0.25°C | ± 0.1°C |
| From 15 to 40°C * | ± 0.15°C | ± 0.1°C |
*Statistically significant data taken at ~15% O2 and 37.5°C. All data
extrapolated therein. Reader to reader precision for any sheath.
** With ambient temperature ~ 25°C
TruFluor is a trademark of Finesse Solutions,
LLC. © Finesse Solutions, LLC, 2010.
TruFluor pH
![]() Features
|
TruFluor pH
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Environmental Specifications | ||||
|---|---|---|---|---|
| Parameter | Units | Min | Typ | Max |
| Operating Temperature | °C | 4 | 45 | |
| Operating Pressure | Bar (psi) | 1(14.7) | 2(29.4) | |
| Storage Temperature | °C | 0 | 22 | 65 |
| Storage Relative Humidity (non-condensing) | % | 0 | 60 | 100 |
| Altitude | feet | 40,000 | ||
| Materials | USP Class VI and 316L electro-polishged SS | |||
| Performance Specifications | ||
|---|---|---|
| |
Min | Max |
| Sample Rate | Sample rate can be set so that samples are taken in increments from 1 second to 30 minutes | |
| Measuring Range (pH) | 5.5 | 8.5 |
| Specification Range (pH) | 5.5 | 7.5 |
| Relative Accuracy (pH) | +/- 0.1 pH units over +/- 0.5 pH range centered at 1-point standardization pH value | |
| Accuracy at 37.5 °C | Depends on accuracy of method and off-line analyzer used for 1-point standardization. | |
| Response Time (sec) | <90 (agitated, 90%) | |
| Drift | < 0.05 pH units over 21 days for a standard process sampling once every 5 seconds | |
| Calibration | Pre-calibrated (gamma-radiation-resistant chip). Conventional 2-point cal possible. 1-point standardization required. | |
| Gamma Radiation | Tested to 50 kGy (typical 25 - 40 kGy) | |
| Temperature | Accuracy | Precision |
| From 4 to 45°C | ± 0.25°C | ± 0.1°C |
| From 15 to 40°C | ± 0.15°C | ± 0.1°C |
TruConnect,TruLogic and TruBio are trademarks of Finesse Solutions, LLC. © Finesse Solutions, LLC, 2009.
TruTorr
![]() Features
|
TruTorr
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Environmental Specifications | ||||
|---|---|---|---|---|
| Parameter | Units | Min | Typ | Max |
| Operating Temperature | °C | 15 | 45 | |
| Operating Pressure | Psi (bar) | 14.7 (1) |
||
| Storage Temperature | °C | 0 | 22 | 65 |
| Storage Relative Humidity (non-condensing) | % | 10 | 60 | 90 |
| Altitude | ft (m) |
10000 (3048) |
||
| Shock tolerance (non-operating) | ship- ping |
JIS Z 0202: 1994 |
||
| Performance Specifications | ||
|---|---|---|
| |
Min | Max |
| Measuring Range (pressure gauge) |
0 psi 0 bar |
7 psi 0.48 bar |
| Drift | 0.2 psig (0.0014 bar gauge) over 21 days | |
| Limit of Detection (psi) | 0.01 | |
| Relative Accuracy (pH) | +/- 0.1 pH units over +/- 0.5 pH range centered at 1-point standardization pH value | |
| Accuracy at 20 °C | Accuracy (at 25°C): ±0.2 psig (±0.0014 bar gauge) or ±2% of the reading, whichever is greater (including drift 21 days) |
|
| Response Time (sec) | ≤ 2 | |
| 90% response (sec) | < 1 | |
| Calibration |
Pre-calibrated 1 point standardization optional through TruBio software |
|
| Cross Sensitivity | None | |
| Temperature | Automatic compensation over operating range | |
TruTorr, TruViu and TruBio are trademarks of Finesse Solutions, LLC. © Finesse Solutions, LLC, 2010.